Reading a ventilator waveform
Ask what is controlled and what is variable. In a constant-flow volume breath, flow is a square and pressure rises. In a pressure breath, pressure is the square and flow decelerates as the lung fills. Volume is the area under the flow curve in either case.
Then ask whether flow returned to zero before the next breath. If expiratory flow is still going when inspiration starts, gas is trapped. A scoop or a beak on a pressure-volume loop is a teaching picture of overdistension, and it is easier to misread than the textbooks admit. When the picture and the person disagree, look at the person and the circuit.